EMFC MK III E.L.F. Magnetic Field Compensation System

To maximize performance:

Make sure detector is aligned with the axes of the room (x,y,z)

Disable all channels. Close door of BOTH EM rooms. Turn off video monitors. Enable all channels.

Plug BNC cable into X channel and observe on oscilloscope. Use small screwdriver to increase gain of X channel until feeback observed on oscilloscope. Reduce gain by 0.5 turns.

Repeat same procedure for Y channel and Z channel.

Do this sequence several times until system converges.